• Broad spectrum

    of products for

    your industry

Cedex Bio HT Analyzer

Bioprocess analyzer

For use in quality control / manufacturing process only.

Cedex Bio HT Analyzer1
expand | collapse
+-
Order information
Cedex Bio HT Analyzer material number and pack size:
Material Number Pack Size
06608116001 1 instrument with PC, monitor,software and accessories
+-
Overview
Automated computerized analyzer for determination of a wide range of analytes in aqueous solutions like cell culture and fermentation media.
Your gains with the Cedex Bio HT Analyzer

Quality
  • Reliable photometric technology based on 40 years of experience
  • High accuracy and precision for entire assay range
  • Broad dynamic range and sensitivity
  • Performance superior to membrane technology
  • 5% variability between instruments and reagent lots
Speed
  • Automated daily maintenance
  • Pre- and post-dilution capabilities
  • Long reagent on-board stability
  • Reagent loading capacity of 32 cassettes
  • 90 samples loaded at one time
  • Average maintenance is just about 100 minutes per year
Value
  • Versatile menu tailored to today’s workload
  • Expandable menu for tomorrow’s requirements
  • One instrument for multiple parameters of mammalian, bacterial, and yeast culture systems
  • Minimal hands-on time: a true walk-away analyzer
  • Consistent results using the Cedex Bio and Cedex Bio HT Bioprocess Analyzers
+-
Specification

System Specifications

Test Throughput (Theoretical Max)
  • Up to 320 tests/hour (including ISE)
Sample Types
  • mammalian cell culture
  • microbial fermentation
Sample Handling:
  • Up to 90 sample tubes in 6 racks on-board, samples can be replaced during operation
  • Cooled rack position for controls and calibrators
  • Automated sample dilution
  • Barcode reading via laser scanner, with immediate STAT recognition
Sample Container Types    
  • Sampling tubes: 5 – 10 mL; 16 x 100; 16 x 75; 13 x 100; 13 x 75 mm
  • Standard cup: 1.5 mL Micro
  • Micro cup: 650 μL
  • Cup on tube definable
Sample Volume    
  • Minimal sample volume of 100 µl for several tests,
  • 2 – 20 µl sample consumption per test
Sample Barcode types  
  • Code 128, Codabar, Code 2 of 5 interleaved, code 39
Reagents Onboard Capacity    
  • Up to 32 Cedex Bio HT reagent cassettes in 8 racks on-board,
  • 50 – 800 tests per cassette,
  • Cooled reagent compartment (10 – 15°C), on-board stability up to 3 months,
  • Calibration typically once per lot
Reaction Cells    
  • Holds 1,000 disposable cuvettes with 5 mm path length and 120 – 240 μL reaction volume
Control Unit    
  • HP workstation running Windows XP, Intel Pentium IV with 512 MB RAM,
  • Dual 40 GB hard drives, CD Rom, floppy drive and built-in modem
System Interfaces  
  • RS 2332 serial interface, bidirectional, built-in modem for remote access
Electrical Requirements
  • 100 – 125/200 – 240 volts AC, 50 or 60 Hz, consumption 1200 VA
Physical Dimensions    
  • Width: 135 cm (53.1 in.)
  • Depth:   66 cm (25.9 in.)
  • Height:  75 cm (28.5 in.)
  • Weight: 320 kg (506 Ibs)
Water Requirements
  • Up to 2 liters per hour in operation mode, Type 1 NCCLS
Certification
  • CE, UL, C-UL